Spectrometric Laboratory

Spectrometric and Laser Laboratory

Spectrometric and laser laboratory assures measurement of spectral parameters in the whole area of visible light (VIS), in ultraviolet area (UV) and near as well as middle infrared area (NIR). Wavelength range is 175nm - 25 μm.
On optical elements and optical assemblies we measure reflectance, transmission and optical density. By spectrometric methods we can also state polarization direction and extinction ratio of polarized prisms or foils.

Given the above mentioned wide range of spectral measurements, it is necessary to use many different types of measuring devices, methods and jigs. In Meopta we designed and produced rotary tables with an incremental sensor for accurate angle set up on measured pieces and also number of jigs for measurement of exacting assemblies.

Equipment:

  • PerkinElmer Lambda 900
  • PerkinElmer Lambda 950
  • PerkinElmer Lambda 1050
  • Nicolet FT IR 6700 for VIS and near IR area 500 – 3000 nm
  • Nicolet FT IR 6700 for near and middle IR area 1 - 25 μm

Spektrofotometr Perkin Elmer

 Specification of parameters and measurement accuracy of PerkinElmer devices

 Measuring Equipment  Configuration  Measured Parameters  Accuracy
 PerkinElmer Lambda 950, 1050  Basic configuration T, OD
AOI = 0°- 80°
λ = 175 - 3300 nm
 ±0,1%
 URA R
AOI = 8°- 68°
190 - 2500 nm
 ±1,0%
 RC tool AR, R
AOI = 6°
λ = 190 - 2500 nm
 ±0,1%
 PerkinElmer Lambda 900, 950  Optical bench AR, R, T, OD
AOI = 0°- 80°
λ = 210 - 2500 nm
 ±0,1-0,5%
 RC tool AR, R
AOI= 6°
λ = 210 - 2500 nm
 ±0,2%
 VW tool R
AOI = 6°
λ = 210 - 2500 nm
 ±0,2%

 

Specification of parameters and measurement accuracy of Nicolet FT-IR 6700 for VIS and IR area

 Measuring Equipment  Configuration  Measured Parameters  Accuracy
 Nicolet FT IR 6700  Basic configuration T, OD
AOI = 0°
λ = 1 - 25 μm
 ±0,5%
 RC tool R
AOI = 8°
λ = 1 - 25 μm
 ±0,5%
 VW tool AR, R
AOI = 6°
λ = 1 - 25 μm
 ±0,5%

 Spektrometr Nicolet FT-IR 6700
 

Laser measurement

Measurement of spectral parameters on one wavelength.

Measurement system includes rotary table (possibility to set up angle of incidence of laser beam on measured piece in range of 0°-80°) and detector connected to power meter.

This system can be upgraded with additional parts – e.g. polarizer or phase plate.

Used lasers:

  • UV: 263 nm, 266 nm
  • VIS: 469 nm, 532 nm, 633 nm, 659 nm
  • NIR: 810 nm, 820 nm, 1064 nm

 Laserové pracoviště

For more detailed information please contact us at meopta@meopta.com.